Read commands

Function

Command Response

Description

Actual pressure

IN_PV_1

XXXX mbar or

Unit according to preselection

XXXX Torr or
XXXX hPa

Actual pumping speed

IN_PV_2

XX.X Hz

Time

IN_PV_3

XX:XX h:m

Time for shut down or time of warmup
(only RC 5 / RC 6)

Increase of pressure

IN_PV_4

xxx/min

Only VACUU•LAN

Time of process

IN_PV_5

XX:XX h:m

Only VACUU•LAN

Preselected

IN_CFG

XXXXX

at controller

0: Remote mode off
1: Remote mode on
0: not used
0: not used
0: not used
0: not used
0: No cooling water valve
1: Cooling water valve
0: VACUU•LAN
1: Continuous pumping
2: Vacuum control
4: RC 5 / RC 6 management

Preselected

IN_CFG_1

XXXXXXXX

at controller

0: Remote mode off
1: Remote mode on
0: mbar
1: Torr
2: hPa
0: External error off
1: External error on (evaluation)
2: Stop (control is stopped)
0: Acoustic signal off
1: Acoustic signal on
0: Isolation valve off
1: Isolation valve on
0: Cooling water valve off
1: Cooling water valve on

Control with 0: Valve

1: Pump
2: VARIO
0: VACUU•LAN
1: Continuous pumping
2: Vacuum control
4: RC 5 / RC 6 management