3.4.3 Analog Trigger

You can choose either CH0, CH1, CH2 or CH3 as the trigger sig-
nal while using external analog trigger source. The trigger level
can be set by software with 8-bit resolution. Please refer to
Table 3-3 for the ideal transfer characteristic.

The trigger conditions for analog triggers are illustrated in
Figure 3-9 and described as follows:

Positive-slope trigger: The trigger event occurs when the
trigger signal (analog input signal) changes from a voltage
that is lower than the specified trigger level to a voltage that
is higher than the specified trigger level.

Negative-slope trigger: The trigger event occurs when the
trigger signal (analog input signal) changes from a voltage
that is higher than the specified trigger level to a voltage that
is lower than the specified trigger level.

Trigger Level

Setting (Hex)

Trigger Voltage

(-1V to +1V Range)

Trigger Voltage
(-0.2V to +0.2V)

0xFF

0.992V

0.1984V

0xFE

0.984V

0.1968V

---

---

---

0x81

0.0078V

1.56mV

0x80

0V

0V

0x7F

-0.0078V

-1.56mV

---

---

---

0x01

-0.992V

-0.1984V

Table 3-3: Ideal Transfer Characteristics for Analog Triggers