Calibrate the scanner if colors appear in areas that should be white,

or colors of the scanned image and the original vary a lot. This allows

you to adjust and compensate for variations that can occur in the optical

component over time by its nature.

The following describes how to calibrate the scanner:

1. The included Calibration Sheet, marked with an arrow on either edge at

both sides, is used specifi cally for calibrating the ImageScan Pro

940u.

2. Insert the special Calibration Sheet into the ADF of the scanner, with the

arrow pointed toward the scanner buttons.

3. Click Start > All Programs > Your scanner model name > Calibration

Wizard.

4. Click Next from the Calibration Wizard that pops up, then click Calibrate

Now....

5. Click on the Start button to begin the calibration process.

6. The scanner pulls the calibration sheet through the scanner, which

completes thecalibration process. After the calibration is complete, click

OK and then click Finish.

Calibration

Cleaning the ADF

1. Open the ADF cover by pulling the lever.

2. Gently wipe the following areas with the cloth dampened with

the cleaner. Be careful NOT to scratch their surfaces.

Clean Feed Roller:

Wipe the Feed Roller from side to side and then rotate it. Repeat

this until its entire surface is cleaned.

Clean Pad Module:

Wipe the Pad Module from top to bottom (in the direction of

arrow shown in below picture). Be careful not to damage the

pick springs of the pad.

3. Wait for cleaned areas to dry completely.

4. Close the ADF cover by pushing it back down until it snaps

back into place.

TM

Have the following cleaning materials ready to clean the scanner:

soft, dry, lint-free cloth

DO NOT USE WATER

non-abrasive cleaner fl uid: Isopropyl

rubbing alcohol(95%)