ENMET SE-5175 User Manual

Page 4

Advertising
background image

SE-5175

ENMET Corporation

2


2.0 Components of the

SE-5175

2.1 SE-5175 elements

See Figure 1 for location of elements:

Feature

Description

Enclosure

A fiberglass-reinforced polyester 5 x 4 x 3, with a detachable front cover.

Four holes for mounting the enclosure to a vertical surface. Located at the corners of the
bottom of the enclosure. See Figure 3

Front Cover

Detachable front cover of

SE-5175

with Display Panel. See Section 2.2 and Figure 1

There are 4 Screws that hold the front cover in place.

2.2 SE-5175 Operational Features

The Display Panel is attached to the enclosure cover and is released by unscrewing the 4 screws located in the corners. After
releasing the screws and lifting the cover and exposing the interior of the enclosure, see section 3.0 for installation and wiring
instructions.
See Figure 1 for location of features.

Feature

Description

Display

A single line, 8 character LCD with backlight. Indicates the level of gas detected by sensor.
The numerical value of gas concentration and other information is displayed.

Visual Indicators and
Alarms

LED indicators:

Power / Fault Indicator LED, Green / Red
Alarm (3) Indicator LED, Red

Membrane Switches

2 Pushbutton Switches on front panel

control the instrument maintenance functions. The

pushbutton switch locations are indicated by:

M

ENU

↓: Advances the instrument display through operation information and maintenance

menus

S

ELECT

: Disables audio alarm temporally and

Selects the maintenance menu operations such as, Zero, Span, Exit menu or sets
proper calibration values for Zero or Span

See Section 4.0 and 5.0 for operational and maintenance flow charts.

Three alarm points are preprogrammed into the

SE-5175

. At each alarm point, an LED on the front panel is activated. These

internal alarm settings are independent of the 4-20mA output alarm values that can be set at a controller.

Figure 1: External SE-5175 Features

Menu

Select

Display

Strain Relief

Visual Indicators:
Alarm1, Alarm2, Alarm3

Sensor

Pushbutton Switches

Visual Indicator

Power/Fault

Front Cover
Retaining Screws
4 places