Load coil detection test – EXFO CableSHARK P3 VF/DSL Cable Qualifier User Manual

Page 136

Advertising
background image

VF / DSL Cable Qualifier

124

8.6..Load Coil Detection Test

The CableSHARK Load Coil Detection Test can detect if there is any load coil along the line and advise the
user to perform an auto or manual TDR test to locate the load coil(s). When running a TDR test, a load coil
on the line acts like an open circuit on the line. When running FR or DMT tests with a line, the test results will
show the severe roll off in frequencies from 4 kHz to some 10 kHz should loading coils exist on the line.

8.6.1 Setting up a Load Coil Detection Test

From the Main Menu, pressing function button F1 (CABLE TESTS) will bring the user to the CABLE TESTS
pop-up menu. Choosing LOAD COIL will invoke the LOAD COIL DETECTION SETUP menu. The sections
that follow will explain the operation as each of the control keys is pressed and their respective screens
appear. Figure 8.6.1A below shows the Load Coil Test Setup Menu.

EDIT /
SELECT

MAIN
MENU

Test Type :

E

F

CHG

LOAD COIL RX

WARNING

Note: telephone cables may have dangerous A.C. or D.C.
voltages on them, therefore caution should be taken when

connecting this test set.

!

SELECT
NEXT

F

IGURE

8.6.1A

L

OAD

C

OIL

D

ETECTION

S

ETUP

M

ENU

The Load Coil Detection Setup menu has several options that appear at the upper part of the screen (see
the previous illustration). The parameters can be modified to customize a Load Coil Detection test. The
Load Coil Detection test setup options are described as follows: