The EC, Resistivity and TDS scales are auto-ranging. The meter automatically
sets the scale with the highest possible resolution.
By pressing

Lock, the auto-ranging feature is disabled and the current

range is frozen on the LCD.

The

“Range: Locked” message is displayed. To restore the auto-ranging

option press

AutoRng functional key again.

The auto-ranging mode is also disabled by selecting a “fixed range” in
the SETUP menu. While in fixed range mode the instrument will display
the readings with the fixed resolution. Maximum 6 digits can be
displayed.

The top of the fixed range is displayed blinking when the reading exceeds
this value.
To disable fixed range mode enter SETUP and select auto-ranging mode.

Note: Auto-ranging is automatically restored if the range is changed, if

the meter is turned off and back on again or if the calibration
mode is entered.