Hanna Instruments HI 3512 User Manual

Page 69

Advertising
background image

69

• For each record:

• The number of the record (3 chars)
• The reading status (1 char): R, O, U
• The unit of the reading (2 chars): it is sent only

for ISE, EC, Resistivity, TDS, Salinity log only

• The calculated reading, with sign, decimal point

and exponent (12 chars)

• The temperature reading, with sign and decimal

point (7 chars)

• The second reading status (1 char): R, O, U; not

available for EC log

• The second unit of the reading (2 chars); it is sent

only for Resistivity, TDS and Salinity log only

• The second reading, with sign and decimal point

(8 chars); not available for EC log

• The logging time (6 chars): it is sent only for

AutoEnd log

• The temperature probe persence (1 char)

LODxPyyy Requests the yyy

th

pH record logged data when x=1

LODxMyyy Requests the yyy

th

mV/Rel mV record logged data when x=1

LODxIyyy Requests the yyy

th

ISE record logged data when x=1

LODxEyyy Requests the yyy

th

EC record logged data when x=2

LODxRyyy Requests the yyy

th

Resistivity record logged data when x=2

LODxTyyy Requests the yyy

th

TDS record logged data when x=2

LODxNyyy Requests the yyy

th

Salinity record logged data when x=2

LODxPALL Requests all pH Log on demand when x=1
LODxMALL Requests all mV/Rel mV Log on demand when x=1
LODxIALL Requests all ISE Log on demand when x=1
LODxEALL Requests all EC Log on demand when x=2
LODxRALL Requests all Resistivity Log on demand when x=2
LODxTALL Requests all TDS Log on demand when x=2
LODxNALL Requests all Salinity Log on demand when x=2

The answer string contains:

• The number of records (3 chars): this is sent only is

“ALL” data is requested

• The log type (2 chars)
• The number of the record (3 chars)

Advertising