Hanna Instruments HI 3512 User Manual

Page 34

Advertising
background image

34

• Press

CAL from EC range to enter calibration mode.

• Leave the probe in the air and press

EC. The instrument will display

the measured EC on the LCD, first expected standard and the
temperature reading.

• If necessary, press the

ARROW keys to select a different standard

value.

• The “ ” tag will blink on the LCD until the reading is stable.

• When the reading is stable and close to the selected buffer,

Confirm

functional key is displayed.

• Press

Confirm to confirm first point.

• The calibrated value and the second expected standard value is then

displayed on the LCD.

• After the first calibration point is confirmed, immerse the EC electrode

into the second standard solution and stir gently. Tap the probe
repeatedly to remove any air bubbles that may be trapped inside
the sleeve. The instrument will automatically detect the standard
after few seconds the reading is stable.

• If necessary, press the

ARROW keys to select a different standard

value.

• The “ ” tag will blink on the LCD until the reading is stable.