Hanna Instruments HI 3512 User Manual

Page 29

Advertising
background image

29

It is recommended to calibrate the instrument frequently, especially if
high accuracy is required.
The ISE range should be recalibrated:
• Whenever the ISE probe or ion charge is changed.
• At least once a day.
• After testing aggressive chemicals.
• When calibration alarm time out is expired-

“CAL DUE” warning

blinks (if feature is enabled).

Follow instructions for the individual electrode.
Due to electrode conditioning time, the electrode must be kept immersed
a few seconds to stabilize. The user will be guided step by step during
calibration with easy to follow messages on the LCD. This will make the
calibration a simple and error-free procedure.

PROCEDURE
Select the proper ISE probe in SETUP menu or select the proper Ion
Charge slope (see SETUP for details, page 40).
Note: If the ISE probe has not been previously calibrated (at least one

point), the

“----” will be displayed.

Pour small quantities of the buffer solutions into clean beakers. If possible,
use plastic beakers to minimize any EMC interferences.

For accurate calibration and to minimize cross-contamination, use two
beakers for each standard solution. One for rinsing the electrode and one
for calibration.

The

HI 3512 permits a 5 point calibration in 6 standard solutions

(0.1, 1, 10, 100, 1000, 10000). It is advised to bracket expected ion
concentration with your standards. For fluoride electrode the 2 ppm
standard is also available.

For accurate ISE measurements it is recommended to perform a multipoint
calibration. At least a two-point calibration is required.

Remove the protective cap from the ISE electrode.

IIIIIS E

S E

S E

S E

S E CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

Advertising