Hanna Instruments HI 83207-01 User Manual

Page 17

Advertising
background image

17

Aluminum

ALUMINUM

SPECIFICATIONS

Range

0.00 to 1.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.02 mg/L ±4% of reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the aluminon method. The reaction between aluminum and reagents

causes a reddish tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93712A-0

Ascorbic acid

1 packet

HI 93712B-0

Aluminon reagent

1 packet

HI 93712C-0

Bleaching powder

1 packet

REAGENT SETS

HI 93712-01 Reagents for 100 tests
HI 93712-03 Reagents for 300 tests
For other accessories see page 65.

MEASUREMENT PROCEDURE

• Select the

Aluminum method using the procedure described in the

Method Selection section (see page 12).

• Fill a graduated beaker with 50 mL of sample.

• Add the content of one packet of HI 93712A-0 Ascorbic acid and mix

until completely dissolved.

• Add the content of one packet of HI 93712B-0 Aluminon reagent

and mix until completely dissolved. This is the sample.

• Fill two cuvettes with 10 mL of sample each (up to the mark).

10 mL

# 1

10 mL

# 2