Hanna Instruments HI 83207-01 User Manual

Page 8

Advertising
background image

8

The instructions listed below should be carefully followed during testing to ensure most accurate results.
• Color or suspended matter in large amounts may cause interference, and should be removed by

treatment with active carbon and filtration.

• Ensure the cuvette is filled correctly: the liquid in the cuvette forms a convexity on the top; the bottom

of this convexity must be at the same level as the 10 mL mark.

COLLECTING AND MEASURING SAMPLES

• In order to measure exactly 0.5 mL of reagent with the 1 mL syringe:

(a) push the plunger completely into the syringe and insert the tip into the solution.
(b) pull the plunger up until the lower edge of the seal is exactly on the 0.0 mL mark.
(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hanging

on the tip of the syringe, if so eliminate them. Then, keeping the syringe in vertical position above

the cuvette, push the plunger down into the syringe until the lower edge of the seal is exactly on

the 0.5 mL mark. Now the exact amount of 0.5 mL has been added to the cuvette, even if the

tip still contains some solution.

USING LIQUID AND POWDER REAGENTS

• Proper use of the dropper:

(a) for reproducible results, tap the dropper on the table for several times and wipe the outside of the

dropper tip with a cloth.

(b) always keep the dropper bottle in a vertical position while dosing the reagent.

TIPS FOR AN ACCURATE MEASUREMENT

(a)

(b)

Advertising
This manual is related to the following products: