Copper HR

COPPER HIGH RANGE

SPECIFICATIONS

Range

0.00 to 5.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.02 mg/L ±4% of reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the

EPA method. The reaction between copper and the bicinchoninate

reagent causes a purple tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93702-0

Bicinchoninate

1 packet

REAGENT SETS

HI 93702-01 Reagents for 100 tests
HI 93702-03 Reagents for 300 tests
For other accessories see page 65.

MEASUREMENT PROCEDURE

• Select the

Copper HR method using the procedure described in the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample (up to the mark)

and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The meter will show “-0.0-”” when the meter is zeroed and ready for

measurement.