• Press Timer and the display will show the countdown prior to zeroing the blank or, alternatively, wait

for two minutes and press Zero. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.

• Remove the cuvette.

• Insert the other cuvette (# 2) with the reacted sample into the

instrument.

• Press Read to start reading. The instrument displays the results in mg/L of fluoride.

Note: For wastewater or seawater samples, before performing measurements, distillation is required.
For most accurate results, use two graduated pipettes to deliver exactly 8 mL of distilled water and
8 mL of sample.

Fluoride

INTERFERENCES

Interferences may be caused by:
Alkalinity (as CaCO

3

) above 5000 mg/L

Aluminum above 0.1 mg/L
Iron, ferric above 10 mg/L
Chloride above 700 mg/L
Phosphate, ortho above 16 mg/L
Sodium hexametaphosphate above 1.0 mg/L
Sulfate above 200 mg/L
Highly colored and turbid samples may require distillation
Highly alkaline samples can be neutralized with nitric acid.