7 bump test result, Bump test result, Result, test – Heath Consultants Personal Surveyor (PS200 Series) User Manual

Page 49: Test result

Advertising
background image

3-5

MANUAL BUMP TEST OPTIONS

Note: The audible, visual and vibrating alarms activate
for 2 seconds only (default setting) when activated during
the bump test.

3.7 BUMP TEST RESULT

After selecting ‘YES’ or ‘NO’ the user is informed of the
overall bump test result as shown in Fig. 3-5.

Fig. 3-5 Bump Test Pass

The bump test gas should now be removed.
The bump test result including date and time will be
automatically datalogged.
When the gas readings fall below their alarm set-points,
or after 60 seconds, the bump test is complete, and the
instrument will automatically return to normal operation.
Should any gas range fail the bump test, the display will
be red and a

symbol will be displayed as shown in

Fig. 3-6.