Labconco service fixture – Labconco Protector ClassMate Laboratory Hoods 69716 User Manual
Page 52
48
Product Service 1-800-522-7658
Chapter 7: Modifying Your Protector ClassMate Fume Hood
If the service line pressure exceeds 40 PSI, it must be
equipped with a pressure regulator to reduce the line
pressure. Please use the appropriate instructions for the
appropriate valves. See Appendix D: Protector
ClassMate Accessories to order.
Labconco Service Fixture
1. Decide the locations (either position 1 or 3) you
want to install the service fixture and outlet fitting
and remove the hole plugs, and access panels. (See
Figure 7-1)
2. Mount the Labconco Valve (P/N 9817000 or
P/N 9823700 – water only) through the hole and
tighten with the supplied nut (P/N 9818000).
Mount the knob (P/N 98187-XX, lens (P/N
9818900), and label (P/N 9825500) underneath the
lens. XX colors as follows:
-00 Argon - Gray
-01 CW - Green
-02 GAS - Blue
-03 AIR - Orange
-04 VAC - Yellow
-05 HW - Red
-06 DW – White
-07 STM – Black
-08 NIT - Brown
3. Mount the Bulkhead Union (P/N 1449500) on the
sidewall through the 3/4" diameter hole and tighten
the supplied nut (P/N 9819001) with a 7/16"
lockwasher (P/N 1910819).
4. Connect one end of the Outlet Service Tube (Part
numbers listed next), to the Bulkhead Union Tube
fitting by pressing the tube all the way into the
fitting, and through the two ferrules turning the
tubing nut hand tight, and then approximately one
more turn using pliers or a 9/16" wrench.
5. Connect the other end of the Outlet Service Tube
(Part numbers listed next) to the Labconco Valve
fitting by pressing the tube all the way into the
fitting and through the two ferrules turning the
tubing nut hand tight, and then approximately one
!