Step 8: Place the right angle manifold P/N 6977000 on

the back of the monitor so that the exit from the
manifold faces the open slot on the right side
vertical support of the fume hood. Secure the
manifold to the monitor with silicone sealant
P/N 1579100. (See Figure 7-5)


Step 9: With the manifold exit facing the open slot

insert the .50" diameter barbed hose stem P/N
1548101 into the manifold and secure with
silicone sealant P/N 1579100.


Step 10: Connect one end of the supplied air hose to the

.50" diameter tube at the 1.50" extension.
Route the hose as necessary to prevent kinks
and bends which can affect the calibration.
Route through open slot. Trim slightly if
required. Firmly connect the other end to the
stem on the rear of the monitor. Air hose is
now complete.


Step 11: Connect the 9VDC power cord to the power

jack on the monitor.


Step 12: Replace the junction box cover plate with the

new junction box cover plate P/N 6985000
that has a .968" square cutout for the outlet
assembly. Snap-in the 115V outlet assembly
P/N 7907802 into the .968" square cutout. (See
Figure 7-6)


Step 13: Connect the male terminals from the 115V

outlet assembly to the female terminals, which
are part of the hood wiring harness. Refer to
the Wiring Diagram in Appendix C and you
will see black to 1C-black and white to 6-red
and green to ground. Power is now supplied to
the 115V outlet.


Step 14: Attach the junction box cover plate with the

115V outlet to the junction box. Plug the
9VDC power supply into the wired 115V
outlet. Power is now supplied to the monitor