PRW serial bench scale user’s manual

SECTION 7 PARAMETERS


The scale has 6 parameters that can be set by the user plus a method of entering the
calibration section.

Press the UNIT key and M+ key together in the weighing mode.

The display will show the first function, "F0 H-L".
Pressing the TARE key will cycle through the other functions.
Pressing ZERO will allow you to set the function. It may be necessary to either use

TARE or set a value using the G/N key to move the active digit and then using the
TARE key to increment a digit, followed by the ZERO key to enter the value. Use

the UNIT key to leave a parameter unchanged.

For example when the display shows “F0 H-L” press the ZERO key to begin.
The display will show “Set Lo”, press the ZERO key to set the low limit, or press the

TARE

to skip to the next parameter, “Set Hi” for setting the high limit.


After pressing the ZERO key to set a limit, use the G/N keys to change the flashing
digit, then use the TARE key to increment the flashing digit. Continue to the next
digit and set it as needed.

When all digits have been set press the ZERO key to store the value. The display
will go back to the parameter just set, i.e. “Set Lo”. Advance to another parameter if
needed or press the UNIT key to return to weighing.

FUNCTION MENU SETTINGS

FUNCTION SUB-FUNCTION

DESCRIPTION

DEFAULT
VALUE

F0 H-L

SEt Lo

Set a value for the Low limit.

000.000

SEt HI

Set a value for the High Limit.

000.000

F1 toL

to CLr

Clears the accumulation memory
without printing the results.

to P-C

prints the Accumulation memory
total and then clears the memory.

to Prt

prints the Accumulation Total, does
not clear the memory.

F2 u nt

Sets the displayed unit, you can
press TARE key to set ON or OFF,
press ZERO key to sure

kilogram,
g,lb,oz,tl,t
h.