PRW serial bench scale user’s manual

-15-

SECTION 10 CALIBRATION

Press UNIT key and M+ key together when normal weighing mode, display shows

“F0 H-L”, press TARE key until display shows “PrOG”, press ZERO key, display
shows “PIN”, You can press G/NUNITZERO key to enter setting mode, press

TARE key to select "p2 CAL", press ZERO key to sure. press TARE key to

select .While it is showing "CAL " press the ZERO key to enter. The display will

show "linear".
1. Normal calibrate
Press TARE key to select "nonlin".then press the ZERO key to enter,the display
shows: "unload

” Remove any weight from the platform. After stable indicator on,

press the ZERO key.

Then the display will show the last calibration weight used. If this is correct ,you can
continue by pressing the ZERO key. If it is not correct use the M+, G/N, TARE
keys to change the calibration weight value. When it is correct press the ZERO key.

Then

display will show “LoAd”. Place the calibration weight on the scale. After

stable, press the ZERO key. Then the calibration has completed.

2. Linear calibrate
Press the ZERO key to enter,the display shows: "ld 0

”, Remove any weight from the

platform. After stable indicator on, press the ZERO key.

Then

display will show “Ld 1”. Place the 1/3 of capacity calibration weight on the

scale. After stable, press the ZERO key.

Then

display will show “Ld 2”. Place the 2/3 of capacity calibration weight on the

scale. After stable, press the ZERO key.

Then

display will show “Ld 3”. Place the full capacity calibration weight on the scale.

After stable, press the ZERO key. Then the calibration has completed.

If the calibration is acceptable the display will return to normal. If an error message
is shown try calibration again as a disturbance may have prevented a successful
calibration.

If the problem persist then contact your dealer.

After calibration the scale should be checked to verify the calibration and linearity is
correct. If necessary repeat calibration, especially be certain the scale is stable
before accepting any weight.