5.3 Vacuum sensor

NOTE

Please refer to the separate operating manual of the vacuum sensor!


In order to protect the vacuum sensor against transport damage, it comes
supplied in its original packaging. Prior to commissioning the freeze-dryer,
the sensor must be installed.


1

Vacuum sensor

2

Electrical connection
of the vacuum sensor

Fig. 15: Position of the vacuum sensor and electrical connection


• Switch the unit off by actuating the mains power switch.
• Take the vacuum sensor out of its original packaging and fasten it to the

connector with a clamping ring (DIN16KF) and a centring ring (both
included in the scope of supply).

• Plug the connector (installed on the unit) onto the vacuum sensor and

hand-tighten the screw on the connector.

NOTE

It is absolutely essential to comply with the manufacturer’s instructions in
the separate operating manual of the vacuum sensor!