7 Malfunctions and error correction

Malfunctions are displayed in the dialog box "Process & equipment
messages" (see chapter 7.2 - "Process and error messages"). An acoustic
signal sounds when an error message is generated.
• Eliminate the source of the problem (see the following chapter).
• Acknowledge the error message.

Pos: 162 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 163 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche 070 -00 10 Allg em eine Stö ru ngen LSCplus Alpha - Ga mm a, Epsilo n 1 -4_2 -4 @ 25 \mo d_1 404 9840 570 04_ 68. docx @ 184 125 @ 2 @ 1

7.1 General malfunctions

Type of error

Possible reason

Correction

No indication on the display

• No power in the mains supply

(see chapter 7.1.1 - "Power

failure").

• Power cord is not plugged in.
• Fuses have tripped.
• The mains power switch is set

to off.

• Check the mains power supply

fuse.

• Plug in the power cord correctly.
• Check the on-site fuses
• Switch mains power switch ON.

The touchpanel does not react at all
or it does not react correctly

• The sensitivity of the touchpanel

is misadjusted.

• Contact the service department

(see chapter 7.3 - "Service

contact")

The password input fails

• The password is not correct.

• Inform the administrator.
• If you have lost the administrator

password: contact the service
department (see chapter 7.3 -

"Service contact")

Insufficient vacuum

• Incorrect connection of the

small flange connection(s).

• Loosen the connection. Place

the centring ring with the inner
sealing ring in a centred manner
between the flange connections
and connect it with the clamping
ring. Ensure that the centring
ring neither slips out of place nor
gets jammed.

• Dirty or damaged lid or door

seal.

• Clean the lid or door seal and

replace it if necessary.

• The ground-in stopper of the

attached drying chamber is not

installed correctly.

• Grease the ground-in stopper

evenly and over the entire
sealing surface with vacuum

grease.

Leakage in the media drain valve

• The media drain valve is soiled

with drying residues or wool

particles from cleaning cloths.

• The O-rings are worn

• Clean the media drain valve (see

chapter 8.1.1.3 - "Aeration
valve")and replace it if

necessary.

• Replace the O-rings.

Leakage in a rubber valve

• The valve is soiled.

• Check the valves individually

(see chapter 7.1.2.4 - "Rubber

valves")