Freeze-dryer EPSILON 1-4 LSCplus
Freeze-dryer EPSILON 2-4 LSCplus

8 Maintenance and service

Version 04/2014, Rev. 1.4 of 09/12/2014 • sb

93

Translation of the original operating manual

Pos: 197 /20 0 Ch rist/3 61 G T-BA L abo r-Pilot (PROJEKTE)/Epsilo n 2 -4 LSC plus Sta nda rd/0 80 Wa rtu ng u nd I nsta ndh altun g/08 0-0 010 -00 10- 003 0 Belü ftun gsven til (n ur E2 -4) @ 2 5\m od_1 404 984 758 394 _68. docx @ 18 612 7 @ 4 @ 1

8.1.1.3 Aeration valve

Contaminants such as product residues may lead to an insufficient vacuum.
In this case, the aeration valve and the media drain valve must be cleaned.
• Switch the freeze-dryer off and disconnect the mains plug.
• Remove the valve core.
• Clean the valve core and the opening with a moist cloth.
• Clean the O-rings and inspect them for any damage. Damaged O-rings

must be replaced.


1

Valve opening

2

Valve core

3

O-rings

Fig. 78: Valve opening and valve core with O-rings

(example, varies depending of the type of freeze-dryer)


• Reinsert the valve core.
• Put the freeze-dryer into operation again.
If the vacuum is still insufficient, the freeze-dryer must be checked by
qualified specialist personnel (see chapter 7.3 - "Service contact").

Pos: 198 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 199 /20 0 Ch rist/3 61 G T-BA L abo r-Pilot (PROJEKTE)/Epsilo n Sta nda rd/0 80 Wa rtu ng u nd I nstan dhal tung /08 0-0 010 -00 10- 004 0 Wä rme tausc her Epsilon @ 25 \mo d_1 404 984 6783 00_ 68. docx @ 18 5903 @ 4 @ 1

8.1.1.4 Heat exchanger (only air cooled freeze-dryers)

A lamellar heat exchanger is used for cooling the refrigerant that is
compressed by the refrigeration unit. This air-cooled heat exchanger is
located at the back of the unit (see chapter 2.1.1 - "Functional and
operating elements").
Dust and dirt impair the cooling effect of the air flow. Dust on the lamellas
prevents the exchange of heat and, thereby, impairs the performance and
power of the refrigeration unit. Strong soiling may cause the unit to fail.
This is why the selected set-up location should be as clean as possible.
• Check the heat exchanger at least once per month for soiling and clean

it if necessary.

• Please contact the Christ service department if you have any queries

(see chapter 7.3 - "Service contact").