Setup Mode (Cont.)

Starrett Altissimo Operation Manual

Page 14

CAT 2000-24

Probe Force Setting: The Probe Force Mode is used to change the
Measuring Force. The default Probe Force Value of 1 is Calibrated
at the factory to 4oz (1.1 Newton). The tool will measure between
approximately 3.5 and 8 oz (1 and 2.2 Newton)

Press the SETUP Key three times and the Probe Force Icon will
light and a Probe Force Value between 0 and 9 will be displayed.
(See Figure 32) Use the F1- F4 keys to adjust the Probe Force
Value and press the F5 key (

) to accept the Value, each unit is

approximately 0.5oz (0.14 Newton). The Force Icon will appear
above the Probe Calibration Icon in the Upper left corner of the
Display.

Note: To reset the Probe Force to the Calibrated Value, press the
Clear key while in the Probe Force Mode, the display will reset to
1, press the F5 Key to accept the value. The Probe Force Icon will
go out.

Display Resolution
: The Display Resolution Mode is used to
adjust the Display Resolution between 0.0001”, 0.0005” and
0.001” (0.002 mm, 0.010 mm and 0.02 mm). The default is
0.0001”.

Press the Setup Key four times and the Resolution Icon will light.
(See Figure 33) Use the F3-F4 Keys to change the resolution and
press the F5 Key to accept the resolution.

Note:
To reset the Display Resolution to the default value, press
the Clear Key while in the Display Resolution Mode, the display
will reset to .0001” (.002 mm), press the F5 Key (

) to accept the

resolution.

Beeper Volume:
The Beeper Volume Mode is used to adjust the
Beeper volume for use in various noise environments.

Press the Setup Key five times and the Beeper Volume Icon will
light and a Value between 0 and 9 will be displayed. (See figure
34) Use the F3-F4 Keys to increase or decrease the volume and
press the F5 Key to accept the value. Each time the value is
changed the Tool will beep at the set volume.

Note:
To reset the Beeper Volume to the default level, press the
Clear Key while in the Display Resolution Mode, the display will
reset to 3. Then press the F5 Key (

) to accept the volume.


Printer
: The Printer Mode is used to set the Serial Data Output to
transmit continuously whenever a new feature is measured or other
function is completed.

Press the Setup Key six times and the Printer Icon will light and a
value will be displayed. (See Figure 35) To turn the printer on
continuously press the F3 Key (

) to set the value to 1, to turn the

printer off press the F4 Key (

) to set the value to 0. Then press the

F5 Key (

) to accept the volume. If the Printer is set to print

continuously the Printer Icon will stay lit after the F5 Key (

) is

pressed.