Care and Maintenance

Starrett Altissimo Operation Manual

Page 20

CAT 2000-24

Cleaning of the Glass Scale and Guide Ways:

Remove Top Cover –
Use 3/16” (4.78 mm) Allen Wrench
to remove the top Cover. (See Figure 39)
Important: Cleaning the Glass Scale is a delicate
operation, therefore use extreme care.









Disconnect Cables –
Disconnect the Display Cable and the
Ground Cable. (See Figure 40)
Remove Belt Lock Knob – Unscrew Belt Lock Knob and
remove. (See Figure 40)
Remove Outer Casing – Carefully lift the Outer casing off
towards the top of the tool in a straight line. Avoid contact
with any parts inside. (See Figure 40)



Clean Glass Scale
– Carefully clean the Glass Scale with a
soft lint free cloth lightly dampened with Isopropyl
Alcohol. Next, coat the Scale by applying a greaseless
lubricant such as LPS

®

1† by saturating a lint-free cloth

then wiping it onto the surface of the glass and letting it
dry. (See Figure 41)

Clean Slide Bearing Tracks and Slide Bearings
Carefully clean the 3 Slide Bearing Tracks and 3 Slide
Bearings outside diameter with a soft cloth or cotton swab
lightly dampened with Isopropyl Alcohol. (See Figure 42)

Clean Guide ways
– Carefully clean the guide ways and 8
guide way bearings located along both sides of the column
with a soft cloth lightly dampened with Isopropyl Alcohol.
(See Figure 42)

Clean Drive Belt and Rollers
– Carefully clean the 2
Rollers located at the top and 2 Rollers located at the
bottom of the column with a soft cloth lightly dampened
with Isopropyl Alcohol. Clean the side of the Drive Belt
that comes in contact with the rollers.

Replace the Cover
– Reconnect the Display and Ground
Cables and replace the Top Cover. Be careful not to over
tighten the screws.



† © 1997 LPS

®

is a registered trademark of LPS