Troubleshooting

Starrett Altissimo Operation Manual

Page 21

CAT 2000-24

Error Messages:
E01 – This message usually results from an over speed condition. If this message appears on display, turn the power
off, wait 5 seconds and turn the power on again. Pressing the UNDO key and then the CLEAR Key may also clear
the error. If this fails to clear the error condition then it may be necessary to perform maintenance on the tool. See
Maintenance Section.

Probe Calibration
Deviation is high - If the Deviation is over .0002” (.004 mm), then check the tightness of the Probe Arm and Probe.
Check the cleanliness of the Probe Tip and Calibration Block. Make sure that the Probe and Belt Locks are fully
unlocked. Then recalibrate the Probe.

Inaccurate Measurements
Poor measurement repeatability - Check the tightness of the Probe Arm and Probe. Make sure that the Probe and
Belt Locks are fully unlocked. Then recalibrate the Probe and reestablish the Datum. Inaccurate Measurement -
Check to see that the proper Datum is selected.

Icons Flashing
If the Probe Icon in the upper left corner of the display is flashing the Probe needs to be calibrated. If the Datum
Icon in the upper left corner is flashing a Datum needs to be established.

Icons Light
If the word “FORCE” appears above the Probe Icon in the upper left corner of the display it indicates that the
Probing Force is set at a value other than the Calibrated Value. See Probe Force Setting in the Setup Mode Section.

If the battery symbol lights in the upper right corner the batteries should be recharged. See Battery Charging in Care
and Maintenance Section.