Warning, Electrodes, Drive-stun backup – Taser X26P User Manual

Page 22

Advertising
background image

22

Chapter 4

Cartridges & ECD Operation

• Low Nerve or Muscle Mass. If the probes impact in an area where there is very little muscle mass (e.g.,

the side of the rib cage), the effectiveness can be significantly diminished.

• Limited Probe Spread. Probe spreads of less than 4 inches (10 cm) (including drive-stun) may result in

little or no effect and become primarily a pain compliance option.

• Wires Break. If a wire breaks (e.g., during a struggle), the current will not flow to the probes and an

additional deployment may be required. Drive-stun may still be available.

WARNING

Do not become over-dependent on the TASER ECD. No force option, including ECDs, is 100% effective

in every situation. Do not deploy the ECD without following your department policies and procedures.

Electrodes

The front of the X26P ECD has two metal electrodes. These electrodes direct the charge to the electrodes on

the cartridge to initiate deployment of the probes. In addition, the electrodes provide the ability to use the

X26P ECD in a “drive-stun” mode as a traditional stun-gun type ECD.

Electrode

Electrode

Drive-Stun Backup

Drive-stun capability is available with or without a TASER cartridge installed. To apply a drive-stun, place the

safety in the up (ARMED) position and pull the trigger. The drive-stun mode is not designed to cause NMI and

generally becomes primarily a pain compliance option. Probe deployment is usually considered more desirable

if NMI is the desired objective, even at close range. Some of the advantages of probe deployment include:

• Drive-stun is only effective while the ECD is in contact with the subject or when pushed against the

subject’s clothing. As soon as the ECD is moved away, the energy being delivered to the subject stops.

Deploying the probes allows the user to separate from the subject while maintaining control.

• Due to automatic reflex actions, most subjects will struggle to separate from the ECD. Each time the ECD

comes back in contact with the subject, another set of marks may be visible on the subject’s skin. Using

the probes allows for one point of discharge.

• If the probes are deployed, even at very close range, the user may drive-stun to another portion of the

body that is further away from the probes, thereby increasing the possibility of inducing NMI.

Advertising