By Set-point value: Measurements starts when a temperature set-point is

reached. With this option, it is also possibletostart measurements above (log

above) or below (log below) a temperature set-point (channel 1). This option is

not valid for alarms that have already been setup.

By <Start/Stop> Button: It starts when the Start/Stop button at the frontal

side of the logger is pressed.

Stop:

Defines logging stop mode: Options are:

Full memory: Loggings can be stored up to the logger full memory capacity

is reached.

Wrap around: Logging never stops. The LogBox will keep on recording the

readings and when the memory is full it will overwrite the oldest record in a

circular or wrap around manner.

After loggings: The logging will stop after the number of readings here

defined.

By date/time: Logging is stopped on user-defined day and time.

Interval:

Defines the interval between readings: hour, minute and second. When the

logging mode is set to the instantaneous the minimum interval is 1 second.

For mean, maximum and minimum values, the shortest interval between

loggings is 10 seconds.

Channel 1: Opens the Channel1 Settings screen, where channel 1 can be set up. This channel is for

temperature measurements.

Tag:

Defines a name for the temperature registers.

Sensor:

Informs the sensor present in channel 1: NTC. It can not be changed. It can not

be disabled.

Unit:

Defines the measurement unit: ºC or ºF.

Limits:

Informs the temperature measurement range. It can not be changed.

Logging Mode: It defines how the value measured will be logged. Options are:

Instantaneous: The value logged will be exactly the value measured. The minimum interval

between measurements is 10 seconds.

Average:

The value logged will be the mean of ten consecutive measurements taken

within a minimum predefined interval of 10 seconds.

Minimum:

The value logged will be the minimum value found in ten consecutive

measurements taken within a minimum interval of 10 seconds.

Maximum:

The value logged will be the maximum value found in ten consecutive

measurements taken within a maximum interval of 10 seconds.

Alarms:

Enables an alarm that is triggered according to user-defined parameters.