1. INTRODUCTION

LogBox-RHT-LCD is an electronic logger of temperature and relative humidity.

There are sensors that measure such data and values (data) obtained are displayed in
an LCD screen and stored in an electronic memory. Data can be later viewed and
analyzed on a PC where they can be opened in tabular and graphical forms. The
humidity output can be configured to display Relative Humidity values or the
temperature value of Dew Point.

The logger is supplied with LogChart-II software, which enables to set up the

logger operation mode. It is also used for visualization of collected data. Other
parameters such as end of measurements, logging interval, etc., are easily defined
through the LogChart-II software.

Measurements can be exported to and opened in other applications, such as

spreadsheet programs.

2. PRESENTATION

Technical Parameters

Measure Range:

Temperature:

-40 °C to 80 °C.

Relative Humidity (RH):

0 to 100 %.

Dew point:

-40°C and 100°C

Accuracy:

Refer to figure Measurement Accuracy.

Note: A measure error can be ruled out using the OFFSET parameter in the Logchart-II software.

Measurement Resolutions:

Temperature:

0.1 °C (14 bits or 16383 levels)

Relative Humidity (RH):

0.1 % (12 bits or 4095 levels)

Memory capacity:

32,000 (32k) logs or 64,000 (64k)

Half for each channel or total when one of the

channels is not enabled.

Measurement Interval:

1 second min.

18 hours max.

Power supply:

3.6 V lithium battery (1/2 AA), built-in

Estimated autonomy:

Higher than 200 days, with daily data reading. Too

frequent logged data readings may shorten battery

life.

Operating temperature:

From –40 °C to 70 °C

Case:

Polycarbonate

Protection:

Suitable for products with protection level IP65.

Electronic module case: IP65; Sensor Case: IP40