5.

FAULT FINDING

Typical problems are shown on the following table.

Symptom

Possible Cause

Service Hint

Pump does not switch

ON or run (Green LED

not lit).

Battery fully discharged or

unserviceable.

Recharge for 14 hours.

Wiring loom disconnected

at some point / Fault on

joint.

Locate & resolder.

Circuit / Keypad failure.

Return for repair.

Pump does not run

(Green LED lit).

Pump or motor seizure.

Replace pump / motor.

Broken wire or worn

brushes.

Repair connection.

Pump runs but no air

drawn.

Diaphragm rubber split.

Replace.

Tubes between pump

and outlet split or

disconnected.

Replace.

Pump runs but cannot

achieve desired flow

rate.

Leak on pump assembly. Locate & Correct.
Dirty or damaged valves.

Clean / Replace.

Inlet filter clogged.

Remove & Clean.

Excessive pressure drop. Check sample filter.

Red error LED lit or

flashing plus audible

alarm.

Battery voltage too low.

Recharge battery.

Excessive pressure drop

or high pump current.

Remove pressure load

/ Repair pump.

Flow rate not main-

tained through full

working shift.

Incomplete charge.

Re-charge.

Battery below capacity.

Fully charge & check

voltagewhich should

be more than 4.8 V. If

below this level, check

the output from the

charger.

Replace battery.

High current consumption

due to mechanical defect

in pump.

Locate & correct.

Filter disc clogged.

Replace.

Fault Finding

Page 12 of 16

VORTEX Standard 2 & VORTEX Timer 2

PERSONAL SAMPLING PUMPS Handbook