An alternative label design is provided should frequent access to the control

be desired.

2.6.1 Adjustment Technique

If the sample flow rate drops due to an increase in pressure loading, adjust

the “feedback” control with a trimming tool and increase the flow to slightly

above the desired flow rate. Reduce the flow back to the desired flow rate

using the front panel keys.

If an increase in the applied pressure loading causes the sample flow rate to

increase, decrease the flow to below the desired flow rate using the

“feedback” control. Increase the flow back to the desired flow rate using the

front panel keys.

Repeat the “feedback” adjustment procedure until the sample flow rate

remains constant over changes in the operating pressure.

An accessory kit is available to provide adjustable inlet pressure loading.

Please contact Casella CEL sales for further information.

2.7

Flow “Shut Down” Adjustment

The flow shut down is factory set to a level ensuring the sampler shuts down

if operated with an inlet blockage or beyond its designed operating envelope.

Adjustment of the shut down threshold may be required for operation at

higher flowrate or pressure drop conditions (for example, when the sampler

is used in conjunction with a, low flow constant pressure regulator).

Operation

Diagram showing the location of

the flow compensation control

potentiometer.

Feedback
Control
(USED
FOR
SETUP ON

Feedback

Control

(Used

during

Setup)

Page 7 of 16

VORTEX Standard 2 & VORTEX Timer 2
PERSONAL SAMPLING PUMPS Handbook