Casella CEL Vortex personal sampling pumps User Manual

Page 8

Advertising
background image

Adjust the shut down threshold value as follows.

1.

Loosen the four screws on the face of the sampler and carefully

remove the lid.

2.

The shut down control is located on the rear of the circuit board as

shown below.

3,

Adjust the potentiometer fully clockwise, until it “clicks”.

4.

Connect the pump via the appropriate sampling head and filter to a

flowmeter, and adjust to the required flow rate using the front panel

keys.

5.

With the pump running and the filter head connected, adjust the

potentiometer anti-clockwise until the “BATTERY/LIMIT” LED

flashes, then back the adjustment off half a turn.

6.

Replace the sampler lid.

2.8

Timer (Timer 2 only)

The LCD timer display situated on the front panel indicates the sample run

elapsed time in whole minutes. The timer can be reset to zero by holding

down the “RESET” key for 3 seconds. This can be carried out while the

pump is on or off.

Operation

BUZZER

FLOW SHUT DOWN
ADJUSTMENT POT
(BLUE)

Diagram showing the location of the flow shut down potentiometer

Buzzer

Flow “Shut Down”

adjustment

potentiometer (blue)

Page 8 of 16

VORTEX Standard 2 & VORTEX Timer 2

PERSONAL SAMPLING PUMPS Handbook

Advertising