Casella CEL Vortex personal sampling pumps User Manual

Page 6

Advertising
background image

Flow Fault Shut Down

The red “BATTERY/LIMIT” LED flashes rapidly and the audible alarm

sounds whenever the pump’s loading exceeds a pre-set value. This

may be due to excessive back-pressure from a tube restriction, filter

blockage, or pump fault.

Should the fault condition remain for 8 seconds, the pump will be

automatically shut down and the “BATTERY/LIMIT” LED remain lit.

Pressing both keys will cancel the error warning and switch the pump off.

Re-start the pump as normal.

2.5

Setting the Flow Rate

Adjust the pump flow to the desired rate while the pump is running.

1.

Press the “FLOW” key for 3 seconds, the green “PUMP ON” LED will

flash.

2.

Temporarily release the key, the green LED will continue to flash

while the pump is in this mode.

3.

Adjust the flow rate, either up or down, by holding down the

appropriate key, the adjustment is slow to start then picks up speed.

4.

If there is no key press after 4 seconds, the pump will return to

normal run mode and the green “PUMP ON” LED will stop flashing.

For optimum stability of flow, it may be necessary to allow the pump to run

for a few minutes before use.

2.6

Flow Compensation

The VORTEX Standard 2 and VORTEX Timer 2 control circuitry is designed to

maintain a stable inlet flow rate over varying pressure drop conditions. Small

increases in back pressure may occur during sampling due to the gradual

increase in filter contamination. All personal samplers are factory calibrated to

suit the majority of monitoring applications and should require no further

adjustments. If however, the sampler is used over a large range of flow rates

or pressure drops, then for optimum stability of a given flow rate, adjustment

of the sampler’s feedback control may be necessary.

The feedback control is located on the side of the pump under a sealed label.

For adjustment of the feedback control it is necessary to remove this label.

Operation

Page 6 of 16

VORTEX Standard 2 & VORTEX Timer 2

PERSONAL SAMPLING PUMPS Handbook

Advertising