ELECTROMAGNETIC COMPATIBILITY INFORMATION

Guidance and manufacturer’s declaration - electromagnetic emissions

The IL 50 is intended for use in the electromagnetic environment specified below.

The customer or the user of the IL 50 should assure that it is used in such an environment.

Emissions

Compliance

Electromagnetic environment - guidance

RF emissions

CISPR 11(EN 55011)

Group 1

The IL50 uses RF energy only for its internal function. Therefore, its RF

emissions are very low and are not likely to cause any interference in

nearby electronic equipment.

RF emissions

CISPR 11 (EN 55011)

Class B

The IL50 is suitable for use in all establishments, including domestic

establishments and those directly connected to the public low-voltage

power supply network that supplies buildings used for domestic pur-

poses.

Harmonic emissions

IEC 61000-3-2

(EN 61000-3-2)

Class A

Voltage fluctuations/

flicker emissions

IEC 61000-3-3

(EN 61000-3-3)

Complies

Guidance and manufacturer’s declaration - electromagnetic immunity

The IL50 is intended for use in the electromagnetic environment specified below. The customer or

the user of the IL50 should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environ-

ment --guidance

Electrostatic

discharge (ESD)

IEC 61000-4-2

(EN 61000-4-2)

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, con-

crete or ceramic tile. If floors

are covered with synthetic

material, the relative humidity

should be at least 30 %.

Electrical fast transient/

burst IEC 61000-4-4

(EN 61000-4-4)

±2 kV for power

supply lines

±2 kV for power

supply lines

Mains power quality should

be that of a typical commer-

cial or hospital environment.

Surge

IEC 61000-4-5

(EN 61000-4-5)

±1 kV line(s) and neutral

±1 kV line(s) and neutral

Mains power quality should

be that of a typical commer-

cial or hospital environment.

Voltage dips, short in-

terruptions and voltage

variations

on power supply

input lines

IEC 61000-4-11

(IEC 61000-4-11)

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5s

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5s

Mains power quality should

be that of a typical commer-

cial or hospital environment.

If a dips or an interruption of

mains power occurs, the cur-

rent of the IL50 may be drop-

ped off from normal level,

it may be necessary to use

uninterruptible power supply

or a battery.

Power frequency

(50/60 Hz) magnetic field

IEC 61000-4-8

(IEC 61000-4-8)

3 A/m

Not applicable

Not applicable

NOTE UT is the a.c. mains voltage prior to application of the test level