Guidance and manufacturer’s declaration – electromagnetic immunity

The IL50 is intended for use in the electromagnetic environment specified below. The customer or

the user of the IL50 should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Conducted RF

IEC 61000-4-6 (EN

61000-4-6)

Radiated RF

IEC 61000-4-3 (EN

61000-4-3)

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2.5 GHz

3 Vrms

3 V/m

Portable and mobile RF communications equip-

ment should be used no closer to any part of

the IL50, including cables, than the recom-

mended separation distance calculated from

the equation applicable to the frequency of the

transmitter.

Recommended separation distance

where P is the maximum output power rating

of the transmitter in watts (W) according to the

transmitter manufacturer and d is the recom-

mended separation

Distance in metres (m).

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site survey,

a

should be less than the compliance level in

each frequency range.

b

Interference may occur in the vicinity of equip-

ment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption

and reflection from structures, objects and people.

a.

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and

land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted

theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an

electromagnetic site survey should be considered. If the measured field strength in the location in which

the IL50 is used exceeds the applicable RF compliance level above, the IL50 should be observed to verify

normal operation. If abnormal performance is observed, additional measures may be necessary, such as

re-orienting or relocating the IL50.

b.

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.