Digi-Star Moisture Tracker User Manual

Page 21

Advertising
background image

21

Moisture Tracker

TM

Operators Manual

D3973 Rev C

Select

Calibrate Sensor press

NOTE: Sensor calibration is

also a main menu option.

• Make sure the six digit Code

number at the bottom right

hand corner of the screen

matches the code number

on the calibration disk. See

page 22 to set code in the

device.

To check Calibration

• Place target calibration disk

directly in front of probe

grate.

• Hold disk steady during cali-

bration, press

1

• If calibration is successful a

pass calibration screen will

display. Device is now ready

for use.

• Check calibration of device

once per day or if air tem-

perature changes +/- 10

o

F

from previous calibration.

• For best accuracy, put the

device outside for 30 minutes

(to acclimate the device) with

power on.

• Keep calibration disk clean

and out of sun.