Short Form Calibration

14

GT400 User’s Manual

D3831 Rev B

SHORT FORM CALIBRATION

The Short Form Setup & Calibration procedure allows you to change “SETUP”

and “CAL” numbers of indicator.

Do not attempt to calibrate scale if indicator is not reading stable weights.

Calibration procedure will not fix instability, inconsistencies, or flashing

"RANGE" messages.

Obtain Current Set-up and Calibration Number

Write down current SETUP and CAL

numbers of your GT 400 indicator.

These numbers are displayed during

Self Test.

SETUP NUMBER

To run self test with indicator ON:
1. Press until LB symbol starts

to blink. Release, then press

to start Self Test.

2. Press to “pause” the Self-Test

while numbers are displayed.

3. Press again to allow self-test

to complete normally.

SETUP # _______________

CAL # ________________

Following is a list of functions that are controlled by the “SETUP” number:

Weigh Method (

W MTHD

) Gain

Display Units (

LB-KG

)

Scale

Capacity

Display Counts (

COUNT

)

1

3

2

1