Reset, Using the calibrator tool – Flowserve SoftTools Suite 7.0 User Manual

Page 10

Advertising
background image

VLAUG001-00 ©Flowserve Corporation, Valtek Control Products, Tel. USA 801 489 8611 10 of 21

Reset

If it becomes desirable to reset the device to the factory default settings, proceed as follows:

From the TechCheck Configurator menu, click on Configurator and select Reset
Configuration.

From the Reset window, you may set all non-volatile memory to factory defaults. Any
custom configurations or settings are lost and the positioner will operate as a standard
model (read the warning within the window). To reset the device, click on the Reset
checkbox and click Apply.

Using the Calibrator Tool

The Calibrator Tool consists of several separate calibration windows and allows calibration
of the actuator pressure sensors, the 4-20 mA current loop, and the closure member stroke
position sensor. Select and perform each of the calibrations to ensure correct operation of
the device. An additional function provides a means to check the device calibration.

1. From the main TechCheck window menu, click on Tools.
2. Select

Use Calibrator. Or,

3. Click on the Calibrator toolbar button for direct access to the main Calibrator

window.

Figure 6 TechCheck Calibrator main menu window

To select a specific calibration type:

1. From the main TechCheck Calibrator window menu, click on Subsystem.
2. Select the type of calibration desired. Or,
3. Click on a toolbar button for direct access to the desired calibration type.

Click on Calibrate Actuator Pressure Sensor to display the Calibrate Actuator Pressure
Sensor
window. This calibration procedure is available only on 121X and 321X (advanced)
models of the Logix positioners, and effects the calibration of the actuator pressure sensors.

Advertising