background image

medSim 300B

Operators Manual

4-28

Intrauterine Pressure (IUP) Simulation

The intrauterine pressure (IUP) waveform is output at "BP4/AUX".
The following drawing illustrates the medSim 300B simulated intrauterine pressure

waveform.

0

Seconds

90

10

90

mmHg

eic198.eps

Figure 4-23. medSim 300B Simulated Intrauterine Pressure Waveform

At the blood pressure "SENSITIVITY" switch located on the front panel, adjust the

medSim 300B to match the input sensitivity of the fetal monitor—either 5 µV/V/mmHg

or 40 µV/V/mmHg.
The frequency of the uterine contractions generated by the medSim 300B can be varied.
To vary the rate at which contractions occur:

1. Select <F1>-ECG from the FUNCTION SELECT MENU for the following two

pages of display:

ECG

BASE

PERF

ARRH

>

<F1>

<F2>

<F3>

<F4>

PVC

eic199.eps

ECG

<

PACE

INS

FETAL

<F1>

<F2>

<F3>

<F4>

eic200.eps

Advertising