Late inflation, Early deflation – Fluke Biomedical medSim 300B User Manual

Page 130

Advertising
background image

medSim 300B

Operators Manual

4-32

Late Inflation

Dicrotic Notch

Inflation

Deflation

Assist Ratio = 1:2

Unassisted
Waveform

Assisted
Waveform

eic207.eps

Figure 4-27. medSim 300B Simulated Late Balloon Inflation

Inflation begins after the dicrotic notch (aortic valve closure).

Deflation ends just before the next arterial pressure waveform.

Note

The dicrotic notch is visible prior to the balloon inflation point.

Early Deflation

Dicrotic Notch

Inflation

Deflation

U Wave

Assist Ratio = 1:2

Unassisted
Waveform

Assisted
Waveform

eic208.eps

Figure 4-28. medSim 30013 Simulated Early Balloon Deflation

Inflation begins at the dicrotic notch (aortic valve closure).

Deflation ends too soon, causing a large U-wave before the next arterial

waveform.