Fluke Biomedical medSim 300B User Manual
Page 75
Operation
Operating the medSim 300B
3
3-47
2. From this menu, press the function key that corresponds to the user sequence that
is to be programmed.
When a sequence choice is made, the display returns to the FUNCTION
SELECT MENU and an asterisk (*) at each end of the second line indicates that
the "make" sequence mode is running. The menu display looks like this:
FUNCTION SELECT
ECG
RESP
BP
TEMP
<F1>
<F2>
<F3>
<F4>
*
*
>
eic112.eps
The following example shows how a waveform segment of a user step-sequence is saved
and placed into a sequence. As an illustration, an ECG baseline BPM setting with its run
time is placed into a user step-sequence.
After a sequence (<F1>-SEQ1, <F2>-SEQ2, <F3>-SEQ3, or <F4>-SEQ4) has been
chosen from the SEQUENCE TO MAKE MENU:
1. Press <F1>-ECG from the FUNCTION SELECT MENU:
ECG
BASE
PERF
ARRH
PVC
<F1>
<F2>
<F3>
<F4>
*
*
>
eic113.eps
2. Press <F1>-BASE:
BASELINE ECG
BPM
AMP
ST
AXIS
<F1>
<F2>
<F3>
<F4>
*
*
>
eic114.eps
3. Press <F1>-BPM:
baseline bpm
30
60
80
120
<F1>
<F2>
<F3>
<F4>
*
*
>
eic115.eps