Parameter overview, Power parameters, Programmed limits – Hoefer PS200HC User Manual

Page 15: Power supply output limits

Advertising
background image

Parameter overview

Power parameters

Because the power output limit is 200 W, and
W = V × A, the maximum current of 2.0 A
can be delivered only if voltage is ≤100 V.
Conversely, the maximum voltage of 200 V can
be delivered only if the current is ≤1.0 A.

Programmed limits

Variable 

Range 

Increment

Voltage

5 – 200 V

1 volt

Current

0.01– 2.00 A

0.01 amp

Time

0 – 999 minutes

1 minute

Timer = 0 = continuous run
Timer = 1 – 999 = timed run in minutes


Power supply output limits

The PS200HC can be programmed to limit volt-
age (V) or current (A). The power supply will
operate at the set limit provided the “load,” or
power requirement, does not exceed the power
supply rating.

“Crossover” occurs when the second parameter
limit is reached as system resistance changes
(Fig 3). At this point, the second parameter
becomes constant, and the previously unchang-
ing param eter adjusts downward as system
resistance changes. (System resistance is affected
by discontinuous buffers, changes in tempera-
ture, and buffer depletion.) When the power
supply is delivering power, the light above the
CONSTANT button indicates which parameter
is being held constant.

p5

Please note the following power
output limitations:

• The maximum current of

2.0 A is available at 100 V
or less.

• The maximum voltage setting

of 200 V is available at 1.0 A
and less.

Note: In most cases, only one
parameter can be held constant
at a time.