Power Failure Detection (PFD)

The Power Failure Detection feature will save
the time, voltage and current settings in the
event of a power outage, and automatically
continue the run when the power is restored.

To alert the user that a power failure occurred
during the run, the error message “E6” flashes at
the end of the run, instead of the usual “OFF”
message. The “E6” error message means that
despite the power outage, the power supply
successfully supplied the programmed voltage
and current for the entire length of time that
was programmed. The time shown at the end of
the run is the time the gel ran since power was
restored, not the total time of the run.

The default settings for Power Failure
Detection are:

Run

Time Setting

PFD

Code

Timed Run

1– 999 minutes

ON

PF1

Continuous run

0

OFF

PF0

The Power Failure Detection feature can be
deactivated during a timed run.

1 Press the SELECT button to move the indicator

light under the Clock symbol.

2 Press the UP and DOWN buttons to enter in the

run time in minutes.

3 Press both the UP and DOWN buttons at the same

time. The display will briefly show “PF0”, indicat-
ing that the PFD mode is disabled.

4 Press both the UP and DOWN buttons at the same

time again. The display will show “PF1”, indicat-
ing that the PFD mode is enabled.