14

OPERATING PROCEDURE

1. Locate and position the POC so that the Fan Inlet Vent and Fan Outlet

Vent are not obstructed. The unit, particularly the air inlet slots
and the air outlet areas, should be positioned in a well-ventilated
space.

2. Power the POC using one of
the four modes described in the
Powering the LifeChoice Activox
POC section.

3. Press the Power Button once to turn the unit on. An audible beep

will sound and the Power LED, the Auto Mode Active-High LED and

the Pulse Setting of 1 LED will flash
green. This indicates that the POC
is powered on, ready for use, and
the settings may now be changed.
Pressing the Power Button for an
extended time after the audible beep
will cause the system to turn off.

NOTE: After powering on the POC,
allow 3 minutes for the unit to reach
stated performance. After initial

power on, the Pulse Setting LED
corresponding to the last flow rate
used will illuminate.

CAUTION: DO NOT place the cannula into nasal passages until the
POC has fully powered on. Breathing through the cannula during
start-up may prevent the system from powering on.

4. There are 2 inhalation sensitivity
modes on the control panel:
Active-High and Sleep-Low which
automatically adjust based on oxygen
demand (e.g. breathing pattern)
requirements. No action is required
to start the delivery of oxygen.