OPERATION

DESCRIPTION

Auto Mode Activated

During operation, either the “Active-High” or “Sleep-

Low” Auto Mode Indicator will illuminate. If “Active-

High” is illuminated it means your breath rate is typical

for an active user. If “Sleep-Low” is illuminated it means

your breath rate is slower than your normal active breath

rate.

Alarm Muted

When activated, the audible alarm signal will be muted

and the Mute LED will illuminate amber. The amber LED

will remain illuminated on the Control Panel until the

Mute button is pressed again or the device is powered

off.

Breath Detected

The selected Pulse Setting LED will flash as you breathe.

This is confirmation the device is delivering a dose of

oxygen with each inhaled breath.

Power Interruption

The device will emit a two second audible tone and all

LEDs will turn off if the AC or DC Power is interrupted or

intentionally unplugged. If the power was not intentionally

removed, check all connections between the unit and the

power source to ensure all are secure and firmly plugged

in. If the Power LED does not illuminate, switch to another

power source (AC or DC) or source of oxygen and contact

your equipment provider.

Powered On

Unit will emit a brief audible tone and the LEDs for Power,

Active-High and the Pulse Setting of 1 will illuminate

green. After the initial power on, the last Pulse Setting

selected will illuminate. Depending on which battery

the unit is operating from, either the INT or EXT LED will

slowly flash.

Setting Selected

The Pulse Setting LED for the selected setting (1, 2 or 3)

will illuminate green.

NORMAL OPERATION INDICATORS

During normal operation, the Control Panel should appear as
described below: