Inova Labs Activox DUO2 User Manual

Page 19

Advertising
background image

Problem

Light Indication

Audible

Alarm

Action

Low

pressure

detected by

the system.

The red Alarm LED light

flashes on and off every .5

seconds

Audible

tone

alternating

on for

2 seconds

and then

off for 5

seconds.

Turn the system off.

Switch to the POC or another

source of oxygen and contact

your equipment provider.

High

Temperature

detected by

the system.

The red Alarm LED light

flashes on and off every .5

seconds

Audible

tone

alternating

on for

2 seconds

and then

off for 1

second.

Turn the system off.

Switch to the POC or another

source of oxygen. Check the

air vents. If there is blockage,

remove it and restart the

system after 15 minutes. If

there is no blockage, contact

your equipment provider.

AC power

interruption

detected by

the unit.

All LEDs off.

Approx.

2 second

audible

tone.

Switch to the POC or another

source of oxygen. Check the

power supply. If there is a

power supply issue, solve

as appropriate. If there is no

power supply issue, contact

your equipment provider.

Circuit

Breaker

tripped

All LEDs off.

Approx.

2 second

audible

tone.

Reset the internal circuit

breaker, by pressing the

Reset Button. If there is a

power supply issue, solve

as appropriate. If there is no

power supply issue, contact

your equipment provider.

17

ALARM/LIGHT

INDICATORS

continued

ALARM INDICATORS

MUTE BUTTON
When the POC is connected to the Stationary Base, the POC mute button can be
pressed to mute the audible tones from the Stationary Base.

*

NOTE: The unit will not alarm for low oxygen purity during the first five minutes of

Activox DUO2 operation to allow the unit to reach full oxygen production.