29

EMC INFORMATION

Medical electrical equipment needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information
provided in the accompanying tables.

The Activox DUO2™ Stationary Base is intended for use in the electromagnetic
environment specified below. The customer or end user of the model

130-

001 should assure that it is used in such an environment.

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity and
Emissions

Immunity Test

IEC 60601

Compliance

Electromagnetic

Test Level

Level

Environment - Guidance

Floors should be wood,

concrete or ceramic tile.

If floors are covered with

synthetic material, the

relative humidity should

be at least 30%.

+/- 2, 4, 8 kV

Air Direct

+/- 2, 4, 8 kV

Contact Direct

+/- 2, 4, 8 kV

Contact Indirect

+/- 2, 4, 8 kV

Air Direct

+/- 2, 4, 8 kV

Contact Direct

+/- 2, 4, 8 kV

Contact Indirect

Electrostatic

Discharge (ESD)

IEC 61000-4-2

Power frequency magnetic

fields should be at levels

characteristic of a typical

commercial or hospital

environment.

3 A/m

3 A/m

Power Frequency

(50/60Hz)

Magnetic Field

IEC 61000-4-8

Mains power quality

should be that of a typical

commercial or hospital

environment.

+/-2 kV for

power

supply lines

+/-2 kV for

power

supply lines

Electrical Fast

Transient/Burst

IEC 61000-4-4

Mains power quality

should be that of a typical

commercial or hospital

environment.

± 0.5 kV, ± 1

kV differential

mode

± 0.5 kV, ± 1 kV,

± 2 kV common

mode

± 0.5 kV, ± 1 kV

differential mode

± 0.5 kV, ± 1 kV,

± 2 kV common

mode

Surge

IEC 61000-4-5

Mains power quality

should be that of a typical

commercial or hospital

environment. If the user of the

model REF 130-002 requires

continued operation during

power mains interruptions,

it is recommended that

the oxygen concentrator

be powered from an

uninterrupted power supply

or battery.

< 5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

< 5% U

T

(>95% dip in U

T

)

for 5 seconds

< 5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

< 5% U

T

(>95% dip in U

T

)

for 5 seconds

Voltage dips, short

interruptions and

voltage variations on

power supply input

lines

IEC 61000-4-11