Digital room sensor, Troubleshooting, Operator interface – Orion System Digital Room v.1 User Manual

Page 7

Advertising
background image

Operator Interface

DIGITAL ROOM SENSOR

7

Troubleshooting

Table 1: Temperature/Resistance for Type III 10K Ohm Thermistor Sensors

Temperature – Resistance – Voltage for

Type III 10K Ohm Thermistor Sensors

Temp

(ºF)

Resistance

(Ohms)

-10

93333

-5

80531

0

69822

5

60552

10

52500

15

45902

20

40147

25

35165

30

30805

35

27140

40

23874

45

21094

50

18655

52

17799

54

16956

56

16164

58

15385

60

14681

62

14014

64

13382

66

12758

68

12191

69

11906

70

11652

Temperature – Resistance – Voltage for

Type III 10K Ohm Thermistor Sensors

Temp

(ºF)

Resistance

(Ohms)

71

11379

72

11136

73

10878

74

10625

75

10398

76

10158

78

9711

80

9302

82

8893

84

8514

86

8153

88

7805

90

7472

95

6716

100

6047

105

5453

110

4923

115

4449

120

4030

125

3656

130

3317

135

3015

140

2743

145

2502

150

2288

Figure 9: Temperature/Resistance Testing

+

-

T

Digital Room Sensor

Back View

9302

MICR

OC

HIP

PIC

24H

J

256G

P206

NOTE:

,

Table 1

For This Test The Sensor Must Be Disconnected From

Its TSDRSC Cable As Shown. The Meter Must Be Set To
Measure Resistance in Ohms. Use

To Determine If The

Sensor Is Reading The Correct Resistance Value For The
Ambient Temperature. This Resistance Value Should Match The
Temperature Value Listed Next To The Resistance Value In The
Table. The Temperature Should Be Measured With A Separate
Accurate Temperature Measuring Device Located In The Area
Where The Sensor Is Currently Located.

Test Points Labeled
“Extern Therm” On
Sensor Circuit Board

+

-

TSDRSC Cable

From

Sensor To The
VAV/Zone or VCM-X
Controller Modular
Sensor Port

Disconnected

Advertising
This manual is related to the following products: