2

Calibration procedure for MKP - PKP

INSTRUMENT CALIBRATION
PROCEDURES

1 GENERAL GUIDELINES FOR CALIBRATION

For an accurate calibration it is necessary to proceed as follows:
a) - The instrument under calibration should be mounted in its case in

order to keep the internal temperature stable.

b) - The ambient temperature should be stable.

Avoid any drift due to air-conditioning or others.

c) - The relative humidity should not exceed 70%.
d) - The instrument must be in ON condition from 20 minutes at least.
e) - Operate, possibly, in an environment with no electromagnetic

disturbances.

f) - During calibration, connect to the instrument one input at a time.
g) - Before to execute each calibration, be sure that the specific

hardware setting has been made (see "Preliminary hardware
setting" paragraph).

For this calibration procedure it is necessary to use calibrators with
the following accuracy and resolution:

ACCURACY
1) For current input: + 0.025% output + 0.0025% range + 0.01

µ

A

2) For voltage input : + 0.005% output + 0.001% range + 5

µ

V

3) For TC input: + 0.005% output + 0.001% range + 5

µ

V

4) For RTD input: + 0.02 % + 0.0025

/decade.

5) For cold junction compensation: better than 0.1 °C

RESOLUTION
1) For current input: 0.5

µ

A

2) For voltage input: 100

µ

V

3) For TC input: 1

µ

V

4) For RTD input: 10 m

5) For cold junction compensation: better than 0.1 °C

2 PRELIMINARY HARDWARE SETTINGS

2.1 HOW TO REMOVE THE INSTRUMENT

FROM ITS CASE

1) Switch off the instrument.
2) Push gently the lock A on the right.
3) While the lock A is maintained out, slide out the right side of the

instrument (see fig. 1).

Fig. 1

4) Push gently the lock C on the left.
5) While the lock C is maintained out, slide out the instrument

(see fig. 2).

Fig. 2

2.2 MAIN INPUT SELECTION

Set J103 (see fig. 3) according to the desired input type as shown in
the following table.

J103

INPUT TYPE

T/C, RTD,CJ

60 mV

5 V

10 V

20 mA

1-2

open

open

close

open

open

3-4

open

open

close

close

open

5-6

open

open

open

open

close

7-8

open

open

open

open

close

5-7

close

close

open

close

open

6-8

close

close

open

open

open

Fig. 3

2.3 AUXILIARY INPUT SELECTION (option)

Set J102 (see fig. 3) according to the desired input type as shown in
the following table.

J102

INPUT TYPE

5 V

10 V

20 mA

1-2

close

open

open

3-4

close

close

open

5-6

open

open

close

7-8

open

open

close

5-7

open

close

open

6-8

open

open

open

A

B

B

D

C

D

1 3 5 7

2 4 6 8

1 3 5 7

2 4 6 8

J102

J103

CPU
card

1 2 3 4

ON DIP

V301

J205