Calibration procedure for mkp - pkp – Super Systems 20PQ Calibration Manual User Manual

Page 7

Advertising
background image

7

Calibration procedure for MKP - PKP

9.3) "

∀.

∀.

∀.

∀.

∀.

" - 10 V AUXILIARY INPUT CHECK

The lower and the middle display show "

∀.

" followed by the measured

value in counts (see fig 6).

The "A.I.10" "Hr" calibration is correct if the indication is equal to "

∀.

3

0000" + 10 counts.

a) Check the zero calibration, by setting the calibrator to 0.000 V, the

read-out must be equal to "

∀.

0 0000" + 10 counts.

b) Check the linearity by setting the calibrator to 5.000 V The read-

out must be "

∀.

1 5000" + 10 counts.

NOTE: when it is desired to use a different check point, the
following formula describes the ratio between the signal input and
the instrument read-out (in counts).

Instrument readout (in counts) =

input value

10 (V)

30000

c) Push FUNC pushbutton, the instrument will go to the next

calibration group.

10) "In.Ct" - CURRENT TRANSFORMER INPUT CALIBRATION

The upper display will show "In.Ct".

10.1) "Lr" - INITIAL SCALE VALUE

The lower display will show "Lr"

a) Made the specific hardware setting as described at paragraph 2.
b) Connect the instrument under test to the calibrator as shown in

Fig. 13.

Fig.13

c) The upper display will show "In.Ct", the lower display will show

"Lr" while "OFF" will appear on the middle display.

d) Set calibrator to 0.00 mA AC.
e) Push

pushbutton, the middle display will change to "On".

f) After a few seconds, start the calibration by pushing FUNC

pushbutton. At the end of this calibration routine, the instrument
will go to the next step.

10.2) "Hr" - FINAL SCALE VALUE

The upper display will show "In.Ct", the lower display will show
"Hr" while "OFF" will appear on the middle display.

a) Set the calibrator to 50.00 mA AC.
b) Push

pushbutton, the middle display will change to "On".

d) After a few seconds, start the calibration by pushing FUNC

pushbutton. At the end of this calibration routine, the instrument
will go to the next step.

10.3) "

∀.

∀.

∀.

∀.

∀.

" - CURRENT TRANSFORMER INPUT CHECK

The lower and the middle display show "

∀.

" followed by the measured

value in counts (see fig 6).

The "In.Ct" "Hr" calibration is correct if the indication is equal to "

∀.

0

1000" + 10 counts.

a) Check the zero calibration, by setting the calibrator to 0.00 mA AC,

the read-out must be equal to "

∀.

0 0000" + 10 counts.

b) Check the linearity by setting the calibrator to 25.00 mA AC. The

read-out must be "

∀.

0 0500" + 10 counts.

NOTE: when it is desired to use a different check point, the
following formula describes the ratio between the signal input and
the instrument read-out (in counts).

Instrument readout (in counts) =

input value

50 (mA AC)

1000

c) Push FUNC pushbutton, the instrument will go to the next

calibration group.

15

14

mA AC

Generator

11) "FEEd" - FEEDBACK INPUT CALIBRATION

The upper display will show "FEEd".

11.1) "Lr" - INITIAL SCALE VALUE

The lower display will show "Lr"

a) Made the specific hardware setting as described at paragraph 2.
b) Two resistors (1 K

Ω,

1/4 W, 1%) must be connected to the

instrument under test as shown in Fig. 14.

Fig.14

c) The upper display will show "FEEd", the lower display will show

"Lr" while "OFF" will appear on the middle display.

d) Made a short circuit between terminal 13 and 14.
e) Push

push-button, the middle display will change to "On".

f) After a few seconds, start the calibration by pushing FUNC

push-button. At the end of this calibration routine, the instrument
will go to the next step.

11.2) "Hr" - FINAL SCALE VALUE

c) The upper display will show "FEEd", the lower display will show

"Hr" while "OFF" will appear on the middle display.

a) Remove the short circuit between terminal 13 and 14. Made a

short circuit between terminal 12 and 13.

b) Push

push-button, the middle display will change to "ON".

d) After a few seconds, start the calibration by pushing FUNC

push-button. At the end of this calibration routine, the instrument
will go to the next step.

11.3) "

∀.

∀.

∀.

∀.

∀.

" - FEEDBACK INPUT CHECK

The lower and the middle display show "

∀.

" followed by the measured

value in counts (see fig 6).

The "FEEd" "Hr" calibration is correct if the indication is equal to "

∀.

0

1000" + 10 counts.

a) Check the zero calibration, by removing the short circuit between

terminal 12 and 13 and making a short circuit between terminal 13
and 14; the read-out must be equal to "

∀.

0 0000" + 10 counts.

b) Push FUNC push-button, the instrument will go to the next

calibration group.

12) "O5.ñA" - OUT 5 CALIBRATION.

The upper display will show "O5.ñA".

12.1) "Lr" - INITIAL SCALE VALUE

a) Connect the instrument under test to the calibrator as shown in

Fig. 15.

Fig. 15

b) The upper display will show "O5.ñA", the lower display will show

"Lr" while the middle display will show the actual zero offset in
counts (a number from 0 to 5000).

c) Using

or

push-button, adjust the instrument output until

0.000 mA +0.005 mA is shown by the calibrator.

d) After a few seconds, start the calibration by pushing FUNC

push-button. At the end of this calibration routine, the instrument
will go to the next step.

12

13

14

1 K

1 K

17

+

_

16

Advertising
This manual is related to the following products: