INSTRUMENT CALIBRATION

PROCEDURES

1 GENERAL GUIDELINES FOR CALIBRATION

For an accurate calibration it is necessary to proceed as follows:

a) - The instrument under calibration should be mounted in its case in

order to keep the internal temperature stable.

b) - The ambient temperature should be stable.

Avoid any drift due to air-conditioning or others.

c) - The relative humidity should not exceed 70%.

d) - The instrument must be in ON condition from 20 minutes at least.

e) - Operate, possibly, in an environment with no electromagnetic

disturbances.

f) - During calibration, connect to the instrument one input at a time.

g) - Before to execute each calibration, be sure that the specific

hardware setting has been made (see "Preliminary hardware

setting" paragraph).

For this calibration procedure it is necessary to use calibrators with

the following accuracy and resolution:

ACCURACY

1) For current input: + 0.025% output + 0.0025% range + 0.01 µA

2) For voltage input : + 0.005% output + 0.001% range + 5 µV

3) For TC input: + 0.005% output + 0.001% range + 5 µV

4) For RTD input: + 0.02 % + 0.0025 Ω/decade.

5) For cold junction compensation: better than 0.1 °C

RESOLUTION

1) For current input: 0.5 µA

2) For voltage input: 100 µV

3) For TC input: 1 µV

4) For RTD input: 10 mΩ

5) For cold junction compensation: better than 0.1 °C

2 PRELIMINARY HARDWARE SETTINGS

INPUT TYPE SELECTION

1) Remove the instrument from its case.

2) Set J1according to the desired input type.

3) Set the dip switch V1to the open condition (see fig. 1).

4) Re-insert the instrument in its case.

1

J

t

u

p

n

I
Type

1-2

3-4

5-6

7-8

9-10

TC-RTD

open

close

open

open

open

60 mV

open

close

open

open

open

5 V

close

open

close

open

open

10 V

open

open

close

open

open

20 mA

open

open

open

close

close

NOTE: the not used jumper can be placed on pin 7-9

3 CALIBRATION PROCEDURE

3.1 FOREWORD

Calibration parameters are divided in groups.

Each group is comprised of two parameters (initial and full scale

values) plus a specific calibration check.

Follows a complete list of the "calibration groups" is reported.

A) TC input

B) Cold junction

C) RTD input

D) Linear mA input

E) Linear 5 V input

F) Linear 10 V input

G) Current transformer input calibration

H) Default calibration data loading

3.2 HOW TO PROCEED

1) Switch on the instrument the display will show "CONF".

2) Push the pushbutton, the upper display will show "CAL" while

the lower display is blank.

3) Push the FUNC pushbutton to display the first calibration group on

the lower display. Depress FUNC pushbutton more times until the

desired calibration group is reached.

GENERAL NOTES ABOUT CALIBRATION PROCEDURE:

1) During calibration procedure, when the initial or full scale value of a

group is selected and the upper display shows "OFF", pushing the

FUNC pushbutton the instrument will jump to the next parameter

for checking

without to modify the previous calibration setting.

In this way it is possible to recalibrate only the desired input or

output.

It is also possibe to make a check of one or more calibration group

without to remake the specific calibration.

Pushing the MAN pushbutton it is possible to scroll back through

the calibration parameters without to memorize the setting.

2) During calibration procedure the serial communication interface

and the time out will be disabled.

A) TC INPUT CALIBRATION

A.1) "tL" - INITIAL SCALE VALUE

The lower display will show "tL" while the upper display will show

"OFF"

1) Make the specific hardware setting as described at paragraph 2.

2) Connect the instrument under test to the calibrator as shown in

Fig. 2.

Fig.2

3) Set calibrator to 0.000 mV.

4) Push ▲ or ▼ pushbutton, the upper display will change to "On".

5) After few seconds, start the calibration by pushing the FUNC

pushbutton. At the end of this calibration routine, the instrument

will go to the next step.

1

_

+

3

Fig. 1

J1

1 3 5 7 9

2 4 6 8 10

V1