A.2) "tH" - FINAL SCALE VALUE

The lower display will show "tH" while "OFF" is shown on the upper

display.

1) Set the calibrator to 60.000 mV (see Fig. 2).

2) Push ▲ or ▼ pushbutton, the upper display will change to "On".

3) After few seconds, start calibration by pushing FUNC

pushbutton. At the end of this calibration routine, the instrument

will go to the next step.

A.3) "t....." - TC INPUT CHECK

The upper display show "t." followed by the measured value in

counts as shown in the following figure:

The "tH" calibration is correct if the indication is equal to "t. 3

0000" + 10 counts.

1) Check the zero calibration, by setting the calibrator to 0.000 mV,

the read-out must be equal to "t. 0 0000" + 10 counts.

2) Check the half scale linearity by setting the calibrator to 30.000

mV. The read-out must be "t. 1 5000" + 10 counts.

NOTE: when it is desired to use a different check point, the

following formula describes the ratio between the signal input

and the instrument read-out (in counts).

Instrument readout (in counts) = input value

60 (mV)

30000

3) Push FUNC pushbutton, the instrument will go to the next

calibration group.

B) COLD JUNCTION CALIBRATION

NOTE: make sure that "tL" and "tH" parameters are correctly

calibrated before to calibrate "rJ" parameter.

B.1) rJ - ACTUAL VALUE

The lower display will show "rJ" while the upper display will show

"OFF"

1) Measure the temperature close to terminals 1 and 3 using a

calibrator, for instance, the MEMOCAL (see Fig. 3).

Fig. 3

2) Wait for a few minutes to allow the temperature stabilisation of

the entire system (sensor, calibrator and instrument).

The first pressure of ▲ or ▼ push-button allows to show on the

upper display a read-out value.

3) Using ▲ or ▼ push-button, set a read-out value equal to the

temperature measured by the measuring device (in C° and

tenths of °C).

4) After a few seconds, start the calibration by pushing FUNC

pushbutton. At the end of this calibration routine, the instrument

will go to the next step.

B.2) "rJ" - COLD JUNCTION COMPENSATION CHECK

The upper display will show "rJ." while the lower display will show

the measured cold junction temperature (in C° and tenths of °C).

1) Make sure that the cold junction temperature measured by the

instrument is equal to the value measured by the measuring

device (MEMOCAL).

2) Push FUNC pushbutton, the instrument will go to the next

calibration group.

C) RTD INPUT CALIBRATION

NOTE: make sure that "tL", "tH" and "rJ" parameters are correctly

calibrated before to calibrate this parameter group.

C.1) "PL" - INITIAL SCALE VALUE

The lower display will show "PL" while the upper display will show

"OFF"

1) Make the specific hardware setting as described at paragraph 2.

2) Make a short circuit among the 1, 3 and 4 terminals as shown in

Fig. 4.

Fig.4

3) Push ▲ or ▼ pushbutton, the upper display will change to "On".

4) After a few seconds, start the calibration by pushing FUNC

pushbutton. At the end of this calibration routine, the instrument

will go to the next step.

5) Remove the short circuit from the 1, 3 and 4 terminals.

C.2) "PH" - FINAL SCALE VALUE

The lower display will show "PH" while the upper display will show

"OFF"

1) Connect the instrument under test to the calibrator as shown

in Fig. 5.

Fig.5

2) Set the calibrator to 375.00 Ω.

3) Push ▲or ▼ pushbutton, the upper display will change to "On".

4) After a few seconds, start the calibration by pushing FUNC

pushbutton. At the end of this calibration routine, the instrument

will go to the next step.

C.3) "P." - RTD INPUT CHECK

The upper and lower display show "P." followed by the measured

value in counts.

3

+

_

1

R

TD

1

3

4

Check symbol

Measured or generated

value (in counts)

1

3

4

Check symbol

Measured or generated

value (in counts)